U S E R M A N U A L
MODEL SERIES: PM3000
PM3100 (shown)
PM3200
PM3500
PM3600
SAVE THESE INSTRUCTIONS
CAUTION
Federal (USA) law restricts this device to sale by or
on the order of a physician.
300 Held Drive Tel: (+001) 610-262-6090
Northampton, PA 18067 USA Fax: (+001) 610-262-6080
ISO 13485 Certified www.precisionmedical.com
Zobrazit stránku 0
1 2 3 4 5 6 ... 11 12

Shrnutí obsahu

Strany 1 - U S E R M A N U A L

U S E R M A N U A LMODEL SERIES: PM3000 PM3100 (shown) PM3200 PM3500 PM3600SAVE THESE INSTRUCTIONSCAUTIONFederal (USA) law restricts this device

Strany 2 - CONTENTS

9DISPOSAL INSTRUCTIONSDispose of the Vacuum Regulator in accordance with the local regulations.Please RecycleWARNINGProduct should be cleaned before b

Strany 3

10LIMITED WARRANTY AND LIMITATION OF LIABILITYPrecision Medical, Inc. warrants that the Medical Vacuum Regulator (the Product) will be free of defects

Strany 4 - SPECIFICATIONS

DECLARATION OF CONFORMITY Precision Medical, Inc 300 Held Drive Northampton PA 18067, USA Emergo Europe (European Office) Molenstraat 15 2513 BH

Strany 5 - OPERATING INSTRUCTIONS

1CONTENTS RECEIVING / INSPECTION ... 2INTENDED USE ...

Strany 6 - PARTS DESCRIPTION

2RECEIVING / INSPECTIONRemove the Precision Medical, Inc. Vacuum Regulator from the packaging and inspect for damage. If there is any damage, DO NOT U

Strany 7 - ASSEMBLY INSTRUCTIONS

3SPECIFICATIONS GAUGE RANGE: *Counterclockwise direction PM3000: 0 - 200 mmHg - Full Vacuum PM3100HV: 0 - 300 mmHg - Full Vacuum*PM3000E: 0 - 200 mm

Strany 8 - MRI Safe

4OPERATING INSTRUCTIONSCAUTIONInspect the Vacuum Regulator for visual damage before use,DO NOT USE if damaged.NOTE: Overflow protection should be used

Strany 9 - MAINTENANCE

5PARTS DESCRIPTIONCAUTIONMissing or illegible labels must be replaced, contact Precision Medical, Inc.49875612Selector KnobRegulator Knob1 QUANTITY -

Strany 10 - TROUBLESHOOTING

6REPAIR KITSAnalog Part# Digital Part#PM3000 / PM3000D Vac Reg RK6000 RK6000DPM3000E / PM3000DE Vac Reg RK6000E RK6000DEPM3000HV / PM3000DHV Vac Reg R

Strany 11 - LIMITATION OF LIABILITY

7VACUUM REGULATOR CLEANING ILLUSTRATION(Cleaning/Decontamination Instructions on page 8)Suction TubingColdDisinfection/Sterilization SolutionPatient P

Strany 12 - DECLARATION OF CONFORMITY

8CLEANING / DECONTAMINATION(As needed)1. Attach a working Vacuum Regulator with a continuous regulated mode to a minimum vacuum source of 15 inHg.2. M